Analytical Method - PFC Testing in Water

Sidebar Image

Analytical Method - PFC Testing in Water

Eurofins Eaton Analytical performs PFC analysis by EPA Method 537 v1.1 which uses Solid-Phase extraction (SPE) with liquid chromatography/tandem mass spectrometry (LC/MS/MS) analysis. Samples are collected in 250 mL plastic bottles fitted with a plastic screw-cap. All sample containers include 5 grams/L of Trizma which serves as a buffering agent and to remove free chlorine that may be present in the sample. In addition to sample containers, Eurofins Eaton Analytical also includes supplies for collecting field reagent blanks (FRBs). The FRBs are prepared by Eurofins Eaton Analytical and include an unpreserved container filled with reagent grade water and an empty preserved container to pour the blank water into in the field at the time a sample is collected. FRBs are required by the method but the number of field blanks to collect (for each site, for each representative sample or not at all) is at the discretion of the customer and/or regulator overseeing the project.

Analytical Citation EPA 537 v1.1
Sample Container (3) 250 mLs poly containers
Preservative Trizma
Holding Time Sample 14 days
Holding Time Extract 28 days
Extraction SPE containing SDVB
Elution Methanol
Analysis LC/MS/MS
Surrogates 13c-PFHxA, 13c-PFDA
Internal Standards 13c-PFOS, 13c-PFOA


In September of 2016, EPA issued a Technical Advisory informing of the possible under-reporting of PFOA when adhering to EPA 537 as originally written and recommending expansion of the integration window to account for branched isomers of PFOA. Many of the PFCs can occur in the environment as both branched and linear isomers. However, only a linear standard is currently available for PFOA. The original EPA-approved method did not account for both types of isomers because doing so requires chromatographic peak integration of samples to be calibrated in the same way as the calibration standard, resulting in a limited integration window.

Eurofins Eaton Analytical now includes branched PFOA isomers in the quantitation of PFOA.  We can quantify and report the following PFCs:

 Analyte CAS #
UCMR3
MRL (ug/L)
Low Level
MRL (ug/L)
Perfluorooctanesulfonic Acid (PFOS) 1763-23-1 0.04 0.0025
Perfluorooctanoic Acid (PFOA) 335-67-1 0.02 0.0025
Perfluoro-n-nonanoic Acid (PFNA) 375-95-1 0.02 0.0025
Perfluoro-1-hexanesulfonic Acid (PFHxS) 355-46-4 0.03 0.0025
Perfluoroheptanoic Acid (PFHpA) 375-85-9 0.01 0.0025
Perfluoro-1-butanesulfonic Acid (PFBS) 375-73-5 0.09 0.0025
Perflurobutanoic (PFBA)1 375-22-4 ----- 0.010
Perfluorodecanoic Acid (PFDA) 335-76-2 ----- 0.0025
Perfluorododecanoic Acid (PFDoDA) 307-55-1 ----- 0.0025
Perfluorohexanoic Acid (PFHxA) 307-24-4 ----- 0.0025
Perfluorotetradecanoic Acid (PFTeDA) 376-06-7 ----- 0.0025
Perfluorotridecanoic Acid (PFTrDA) 72629-94-8 ----- 0.0025
Perfluoroundecanoic Acid (PFUnDA) 2058-94-8 ----- 0.0025
N-ethylperfluorooctanesulfonamidoacetic Acid (NEtFOSAA) 2991-50-6 ----- 0.004
N-methylperfluorooctanesulfonamidoacetic Acid (NMeFOSAA) 2355-31-9 ----- 0.004

 